Actual | YOCON automatic nucleic acid extraction workstation can effectively prevent cross-contamination!

Yocon Market

2021-10-20

Company Events


“Accurate, fast, convenient and secure”

This is a prioritization of our requirements for the workstation.

The workstation has to be put into operation.

One of the biggest concerns that customers often have about such product is that -

Whether the equipment will produce a false positive result (cross-contamination of different samples)?

To make customers more assured, we have tested the FE4800 automatic nucleic acid extraction workstation, using real virus samples (purchased from national institutions through regular channels, without infectivity), and have made the first success!


I. Background

What is the possible cause for cross-contamination of samples?

--Too short distance between sampling tubes and between different deep-well plates, and inevitable aerosol produced during liquid transfer by equipment. Cross-contamination of samples may occur if aerosols from one sample fall into another. Our equipment processes 8 samples simultaneously, and it is normal for customers to have doubts.


Why should we prevent such cross-contamination?

--Not all cross-contaminations have serious consequences, but cross-contamination in the molecular biology does produce extremely serious consequences because the back-end testing principle is multiple replication of the initial sample. Based on the current CT value of 35 as the cut-off value for a positive or negative diagnosis during the clinical test, the initial sample should be replicated 235 times, that is, 64.0 billion times. It is amazing that one nucleic acid can be amplified to 64.0 billion nucleic acids. Therefore, whenever a sample is contaminated by another one, result errors are inevitable.


2.Objective

To systematically evaluate the performance of YOCON automatic sample processing and nucleic acid extraction workstation and to verify that there is no aerosol contamination during normal extraction by the equipment.


3.Design

1、 A large number of strongly positive samples and negative samples were arranged in a staggered manner for replicate tests of 3 batches to test whether the extraction equipment would lead to cross-contamination under extreme conditions.

2、 A total of 48 samples were extracted for one batch, including 27 virus-containing samples and 21 virus-free samples.

3、 The virus concentration of the positive samples was 109 copies/mL (refer to the upper limit of actual concentration of positive clinical samples).

4、 Virus-containing samples and blank samples were arranged in a staggered manner, with positive samples around negative sample (negative-6 sample as shown below) and negative samples around positive sample (2-1 sample as shown below), as detailed below:


Arrangement of negative and positive samples.

Negative

Negative

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Negative

Negative

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Negative

Negative

Negative


4.Procedures

1、 The virus mixture was prepared.

2、 The virus samples were added. The virus mixture was added to the sampling tube in a 1:9 volume ratio. A total of 27 virus-containing sampling tubes were set, marked with serial numbers, and recorded. One swab was broken and put into each sampling tube. The tubes were shaken well and mixed well.

3、 Preparation of the reagents required for the instrument. A total of three reagents were required, that is, 15.36 mL of a mixture of isopropanol and magnetic beads (containing 14.4 mL of isopropanol and 960 μL of magnetic beads), 38.4 mL of purification solution and 4.8 mL of cleaning solution.

4、 27 tubes were inserted into the pre-set wells from left to right and top to bottom, and the negative control tube without virus sample were inserted into the remaining 21 wells to start the program for extraction.

5、 The reverse transcription fluorescence quantitative PCR assay was carried out for 21 negative samples, and one well containing viral nucleic acid was set as positive control.

6、 The replicate tests were carried out for 3 batches.


5.Results

1. The expansion curves of each group for batch 1 are as follows:

擴(kuò)增曲線1.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

27.95

Negative control

NoCt




2. The expansion curves of each group for batch 2 are as follows:

擴(kuò)增曲線2.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

28.55

Negative control

NoCt




3. The expansion curves of each group for batch 3 are as follows:

擴(kuò)增曲線3‘’.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

29.72

Negative control

NoCt



6.Conclusions

During the extraction tests for the same batch, negative samples and positive samples were arranged in a staggered manner, and contamination was not found in negative sample wells, but found in positive control well, indicating that this equipment can effectively prevent inter-sample cross-contamination.


Words in the end:

We choose to be one of the few automatic nucleic acid workstation suppliers to offer measured data of samples using our automatic nucleic acid workstation. We will continue to test multiple batches of samples in the future to ensure that the equipment meets the customer needs. Let’s look forward to it!


人妻洗澡被强公日日视频无码动漫 | 精品一区二区三区呻吟声 | 熟女如虎的丰满熟妇啪啪 | 100大看免费行情的软件下载 | 国产精品精品久久久久久 | 精品一区二区三区无码 | 蜜桃成人无码AV在线观看一电影 | 欧洲美女淫妓A级高清视频播放 | 国产一级a毛一级a爰片 | 免费观看黄色视频网站 | 潘金莲的粉嫩小泬12P | 久久精品无码一区二区国产26p | 91在线无码精品秘 黑桃 | 欧美做爰又大又粗XXXⅩ | 日本少妇被强伦轩视频 | 中文字幕日本被黑人无码 | 国产区视频免费在线观看 | 精品国产99久久久久久宅男i | 熟女系列—91Porn | 91网站永久免费看 | 岳乱一区二区三区免费看 | 日本区欧美区亚洲区 | 一级做a视频在线观看 | 色欲17c人妻精品偷拍 | 免费观看BBB毛片大全 | 美女裸体一区二区三区 | 青娱乐性爱视频网站 | 99在线无码精品秘 入口爱酱 | 99人妻字幕乱码久久99一三区 | 果酱短视频成人版黄A片 | 污网站高清无码在线观看 | 精品一区二区三区蜜桃臀绯色 | 1024人妻一区二区三区 | 欧美一区二区三区在线视频 | 欧一美一色一伦一区二区三区 | 熟女人妻 人妻の视频 | 国产91精品秘 入福利姬 | 国产伦子伦对白在线播放观看 | 亚洲AV无码乱码在线观看性色 | 2025中文字幕无码视频 | 鲁鲁视频国产在线播放 |